Dandong Bettersize Instruments Ltd.
|
Payment Terms: | T/T |
Place of Origin: | Liaoning, China (Mainland) |
Add to My Favorites | |
HiSupplier Escrow |
Laser Particle Size Analyzer (Dry & Wet Dispersions)
The Bettersizer 2600 utilizes this laser diffraction technology. There are 92 photoelectric detectors to convert light signals from the scattering spectrum to electrical signals, which are transmitted into an intelligent software. By implementing the Mie scattering theory to the data and performing mathematical conversion, the complete particle size distribution can be derived.
Measurement: Particle size
Particle size range:
Wet dispersion: 0.02 to 2600μm
Dry dispersion: 0.1 to 2600μm
Dispersion type: Wet and Dry
Technology: Laser diffraction
Precise measurement of both large and small particles covering the broadest range by a new patented technology utilizing both Fourier and Reverse Fourier optical systems
Optical bench with 92 detectors covering an angular range of 0.016° - 165°
Optional wet and/or dry dispersion
Small volume dry dispersing module for small sample quantities especially for pharmaceutical or valuable samples in short supply
Easy and fast to changeover between dispersing modules
Easy to implement, create and use Standard Operation Procedures for new materials
Automatic alignment of the optical system
User-friendly software, easily learnt
The Fundamentals of Bettersizer 2600
Sample Dispersion Modules BT-802 Wet Small Volume BT-80N BT-902 BT-903
Sample Dispersion Wet sample dispersion Wet small volume sample dispersion Anti-corrosive sample dispersion Dry sample dispersion Dry small volume sample dispersion
Dispersion Modules
Combine with Main Device
Volume 600mL (dispersion medium) 8 mL (dispersion medium) 80 mL (dispersion medium) 0.2 - 10 g (sample) 0.02 - 1 g (sample)
Automation Full-automated Semi-automated Semi-automated Full automated Full automated
Application Notes
Overview
Features
1)FOURIER AND REVERSE FOURIER OPTICAL SYSTEM
The Bettersizer 2600 is superior in the combination of Fourier and inverse Fourier design. This smart design allows, on the one hand, the detection of stray light in a very wide angular range from 0.016° to 165°. On the other hand, in contrast to the inverse Fourier design, the particles do not have to lie on one plane – the simultaneously accurate measurement of small and large particles is therefore achieved.
Features of combination of Fourier and inverse Fourier design
Spherical detector array: Forward, lateral and backward detectors in 92 pieces in total;
Super large Fourier lens;
Small footprint: Compact design to save space
2)INTUITIVE AND POWERFUL SOFTWARE
Intuitive operation and orderly user interface
Real-time mode for determining the optimal measurement conditions
Fully automatic measurement routine
Automatic cleaning routine
Automatic data backup and highly customizable reports
Direct conversion according to Fraunhofer and Mie evaluation models
Switch between wet and dry dispersion modules with one click
3)WET DISPERSION
BT-802 is designed for particle size measurement with water as the medium.
BT-802 is made up of ABS shell. The components of it include centrifugal pump, peristaltic pump, ultrasonic disperser, pinch valve, control circuit, etc. The stirring speed is adjustable from 500 - 2500 rpm. Built-in ultrasonic system (50 W) ensures stable dispersion of samples.
BT-80N is designed for particle size measurements with organic solvents as the media. BT-80N is suitable for common organic solvents, for example:
BT-80N is made up of stainlesssteel shell. The components of it include centrifugal pump, ultrasonic disperser, PTFE pipeline, sintered-quartz sample cell, control circuit, etc.
Wet Small Volume Dispersion Module is designed for valuable or small-volume sample measurements, where the medium is solvent or water. The module consists of ABS shell, stirring motor, cuvette (8 ml), stirrer, etc.
Maximum volume in 8 mL with samples mass in 0.005 - 0.1 g.
Suitable for samples dispersed in water or organic phase.
4)DRY DISPERSION
BT-902 is suitable for dry powders particle size measurements. The gas used can be compressed air, nitrogen or other noble is.
BT-902 is made up of electromagnetic vibration feeder, venturi pipe, gas circuit, electric circuit, pressure sensor, etc.
BT-903 is designed for particle size measurements of valuable and small amount dry powders with a minimum samples volume of 20 mg.
BT-903 is composed of venturi pipe, gas circuit, electric circuit, samples tube, etc.
5)APPLICATIONS OF BETTERSIZER 2600 PARTICLE SIZE ANALYSER
Performance
Resolution and sensitivity
The resolution and sensitivity of Bettersizer 2600 laser granulometry as verified as below: sample A was gradually added to sample B, the graphs change correspondingly.
Sample A < Sample B
Sample A = Sample B
Sample A > Sample B
SOP
Standard Operating Procedure (SOP)
SOP settings include water intake, bubble removal, background and obscuration measurement, testing, cleaning, and result saving and printing.
Advantages
The Bettersizer 2600 offers an easy solution for standardized and automatic testing.
It ensures the measurement results operator-independent, making the results more objective and reliable.
Technology
Specification
Bettersizer 2600
Parameters measured Materials
Particle size distribution Suspensions, emulsions, dry powders
General
Principle Laser light scattering
Analysis Mie and Fraunhofer scattering
Typical measurement time Less than 10 seconds
Measurement performance
Measuring range 0.02 - 2600 μm (wet) *
0.1 - 2600 μm (dry) *
Accuracy error ≤ 0.5% *
Repeatability ≤ 0.5% *
Reproducibility ≤ 0.5% *
Number of size classes 100 (user adjustable)
Feeding mode Automatic circulation or micro cuvette (wet)
Gas transportation (dry)
Special functions SOP settings, refractive index measurement, sample ratio calculation
Main device
Optical system Combined Fourier and inverse Fourier & Tilted sample cell
Laser High-power optical fiber laser (10 mW / 635 nm)
Detector 92 detectors (forward, lateral and backward arrangements)
Measuring angle 0.016 - 165°
Wet dispersion module
Measurement method One-click operation (automated measurement, cleaning, saving, printing, etc.)
Dispersion medium Water or organic solvents (sample dependent)
Circulation speed 500-2500 r/min
Circulation flow 3000 – 8000 ml/min
Ultrasonication Dry burning prevention,50 W
Dry dispersion module
Dispersion medium Air/ Nitrogen/Noble gas
Air pressure 0.1 - 0.8 MPa (Air compressor dependent)
Gas flow 400 – 6000 L/min
Software
Conformity 21 CFR Part 11, ISO 13320, USP <429>, CE
Reports Customizable reporting
System parameters
Dimensions (L x W x H) 70.5 x 31.8 x 29.5 cm
Weight 23 kg
Supply voltage 100/240V, 50/60 Hz
Computer configuration (recommended)
Computer interface At least one high speed USB 2.0 or USB 3.0 port required
Operating system Windows 7 (32 bit and 64 bit) or higher
Hardware specification Intel Core i5 Processor, 4GB RAM, 250GB HD, Wide screen monitor
*Sample and sample preparation dependent
At Bettersize, our mission is to provide best-in-class particle sizing and characterization instruments, comprehensive solutions and exceptional customer services. We provide particle measuring services, particle size instrument and etc. Want to know particle sizer priceor more? Please contact us.